ESU-2000A (LED) Electrosurgical Unit
ARI-2200 Digital Colposcope (Complete Set)-Digital Colposcope, Camera System, Light Source
Product Details
Features
- Max 300W Electrosurgical Generator, with mono-polar and bipolar function.
- Seven working modes: pure cut, blend1, blend2, contact coag, forced coag, bipolar coag, bipolar cut.
- The electrosurgical generator is used to control bleeding and surgical cutting in all general and open surgical procedures
- Control panel, three-way power output, with different audible and visual indication during the process of outputting.
- Patient return electrode monitoring system and Power Peak System that provides surgeons with improved performance at Lower power settings, minimizing the risk of tissue damage.
- Applicable for Prostate Vaporizing Surgical Operation (TUR), perfectly used under water. The electrosurgical generator can be cooperated with argon plasma coagulator, endoscope and laparoscope.
- Pencil Remoting Output. The output power can be controlled by the remoting control pencil.
- Electrosurgical unit is made of aluminum case with sprayed paint
Configuration
Electrosurgical Pencil 5pcs
Electrosurgical Pad 10pcs
Electrosurgical pad cable 1pc
Footswitch 1set
Bipolar Forceps and cable 1set
Technical Specifications
| PARAMETER |
SPECIFICATION |
| Power |
220V±22V, 50Hz±1Hz (110V±11V, 60Hz) |
| Operating Frequency |
416KHz |
| Power Rating |
1100VA±10% |
| Seven Working Modes |
Monopolar Cut |
a) Pure cut: 1W~300W (Load 800Ω)
b) Blend1: 1W~250W (Load 800Ω)
c) Blend2: 1W~200W (Load 800Ω) |
| Monopolar Coag |
d) Spray Coag: 1W~80W (Load 800Ω)
e) Soft Coag: 1W~120W (Load 800Ω) |
| Bipolar |
f) Bipolar Coagulation: 1W~50W (Load 200Ω) |
| Power Consumption |
≤1100VA |